The glow-wire test answers a narrow but important question: if a hot metal part touches a plastic part, does the plastic sustain a flame long enough to spread fire? It is not a general "is this material flame retardant" test, and treating it as one is the most common reason results get misread. IEC 60695-2-10 defines the apparatus and the procedure; the product standard decides which temperature applies to which part.
What the apparatus actually controls
The glow wire is a nickel-chromium loop with a defined tip geometry, heated by current and monitored by a thermocouple inside the tip. Three things must be right before the first specimen is tested:
- Tip geometry. The loop tip is nominally 4 mm diameter. A deformed or re-ground tip changes the contact area and therefore the heat transfer into the specimen.
- Temperature stability. The tip is held at the test temperature for at least 60 s before application, measured at the tip, not at the power supply display.
- Applied force. The specimen is pressed against the tip with a force of 1 N. Friction in the carriage or a stiff specimen holder will quietly reduce the real force.
GWFI and GWIT are different endpoints
Two indices are quoted from this apparatus, and they cannot be substituted for one another:
| GWFI (Glow-Wire Flammability Index) | Highest temperature at which the specimen extinguishes within 30 s after the tip is removed and does not ignite the tissue paper underneath. The 30 s dwell is the standard application time when qualifying the index. |
| GWIT (Glow-Wire Ignition Temperature) | Temperature 25 K above the highest temperature at which no ignition occurs during a 30 s application. Ignition means a flame that persists longer than 5 s. |
A material can have a respectable GWFI and a modest GWIT. If the product standard asks for "no ignition", a GWFI value is not evidence for it.
Typical test temperatures and where they come from
The apparatus is usually calibrated at 550, 650, 750, 850 and 960 degrees C, and the product standard selects one. Unattended appliances, parts that may be touched, and parts that hold current-carrying conductors are treated differently. Two rules save a lot of argument:
- Test the part as it is mounted. A specimen cut from the moulding and tested flat can behave differently from the same material in a thin rib under a terminal.
- Record the actual tip temperature at the moment of contact. A 15 K drop during application is enough to move a marginal result from fail to pass.
Reading the result: the four observations
Once the tip is withdrawn, four things are recorded, not one:
- Time to extinguish after withdrawal (relevant to GWFI).
- Whether the tissue paper ignites (a drop of burning material is a fail even if the specimen self-extinguishes).
- Depth of penetration of the tip into the specimen.
- Post-test condition: hole, char, cracks, and whether the part still performs its function.
The functional check is often skipped and is often the real finding. A part that passes the flame criteria but has a hole through a creepage path has failed in service terms.
Where results go wrong
- Specimen too small. Heat escapes into the support. Use the specified support and, where the standard requires it, wrap the specimen in aluminium foil.
- Draft. An open window or an extractor hood changes afterflame time more than most operators expect.
- Reused specimens. A previously tested area has a different thermal history and a different surface.
- Temperature read too early. Thermocouple response lags; wait for stability, not for the display to look steady.
A short setup checklist
Before each session: verify tip diameter and condition, confirm the thermocouple is seated in the tip, check the 1 N force with a balance, set the dwell timer to 30 s, lay fresh tissue paper, and confirm the extraction is off. After each specimen: photograph the contact point, record all four observations, and quarantine the specimen until the report is signed.
For equipment selection, the flammability testing range covers glow-wire and needle-flame apparatus; the main catalogue lists the accessory jigs that keep specimen mounting repeatable.